Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System(PCH)

消化管病原体多重核酸検出システム この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

米国回収

24 件

直近 2025.12.10

510(k)

38 件

PMA

0 件

不具合報告

257 件

分類情報

Product Code PCH
デバイス名 Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System

消化管病原体多重核酸検出システム この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

デバイスクラス クラス II
21 CFR 規制番号 866.3990
審査区分 510(k)(市販前届出)
医療専門分野 微生物学
定義 A gastrointestinal pathogen panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and/or symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings.

消化管病原体パネル多重核酸アッセイとは、ヒト便検体から抽出された複数の消化管微生物核酸を同時に検出し識別することを目的とした、定性的なインビトロ診断用医療機器である。消化管感染症の徴候及び/又は症状を示す個人から特定の消化管微生物核酸を検出及び識別することは、臨床評価及びその他の検査所見と併せて用いられる場合、消化管感染症の診断の一助となる。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。

フラグ
該当なし
AI分析

本機器は米国においてクラスIIに分類されており、市販前届出(510(k))による審査区分の対象となっています。510(k)は、既存の同等な機器(プレディケイトデバイス)との実質的同等性を示すことで市販が認められる審査区分であり、クラスIIIに適用される市販前承認(PMA)に比べて相対的にリスクが低い機器に適用されます。

米国市場での動向

本製品コードでは510(k)の累計届出が約40件、申請企業数は16社で、直近5年間も約20件の届出があり継続的な動きが見られます。不具合報告は累計約260件で、その大半が不具合(Malfunction)に分類され、死亡は2件、傷害は約6件報告されており、主な問題としてFalse Positive Result(偽陽性)、Incorrect, Inadequate or Imprecise Result or Readings(不正確・不適切な結果)、False Negative Result(偽陰性)が挙げられています。回収は累計24件で、いずれもクラスII(中程度)に分類され、クラスIの回収は確認されていません。直近5年では13件の回収があり、最新の回収は2025年11月に開始されています。

この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。

関連 product code

同じ 21 CFR 866.3990 に分類される product code(2 件)

No.product code機器名クラス専門分野510(k)PMA不具合回収
1PCI

Gastrointestinal Bacterial Panel Multiplex Nucleic Acid-Based Assay System

消化管細菌核酸同時検出用マルチプレックス核酸検査システム

II微生物学5—50—
2PIQ

Reagents For Detection Of Norovirus Nucleic Acid

ノロウイルス核酸検出試薬

II微生物学1———

機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。

類似の国内医療機器(自動推定)

FDA デバイス分類と JMDN 英語名の名称・定義文の類似度から機械的に推定した候補です。日米の当局が定めた公的な対応関係ではなく、参考情報としてご覧ください。

No.類別一般的名称品目数企業数年間総出荷額成長率平均価格国産比率
1体液検査用器具微生物分類同定分析装置クラスⅠ271014.55 億円/年+23.8% 1,811.46 万円/個0.0%

FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。

米国での回収・承認・不具合報告

無料会員 1,000名 突破

回収・承認・不具合報告の明細の閲覧は会員限定です

無料で会員登録する
回収番号クラス回収開始日企業理由ステータス
Z-0793-2026 PCHクラス II2025.11.03GET TESTED INTERNATIONAL AB

Distribution without premarket approval/clearance.

対応中
Z-0785-2026 PCHクラス II2025.11.03GET TESTED INTERNATIONAL AB

Distribution without premarket approval/clearance.

対応中
Z-0765-2026 PCHクラス II2025.11.03GET TESTED INTERNATIONAL AB

Distribution without premarket approval/clearance.

対応中
Z-0764-2026 PCHクラス II2025.11.03GET TESTED INTERNATIONAL AB

Distribution without premarket approval/clearance.

対応中
Z-0763-2026 PCHクラス II2025.11.03GET TESTED INTERNATIONAL AB

Distribution without premarket approval/clearance.

対応中
Z-2514-2024 PCHクラス II2024.05.16BioFire Diagnostics, LLC

Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.

対応中
Z-1140-2024 PCHクラス II2024.01.26BioFire Diagnostics, LLC

Due to potential signals of increased false positive Norovirus results when using the gastrointestinal (GI) panel.

対応中
Z-0703-2024 PCHクラス II2023.11.27BioFire Diagnostics, LLC

A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.

対応中
Z-2023-2023 PCHクラス II2023.06.05Luminex Corporation

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.

対応中
Z-2588-2023 PCHクラス II2023.04.25BioFire Diagnostics, LLC

The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel

対応中
Z-1392-2023 PCHクラス II2023.03.08BioFire Diagnostics, LLC

Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results.

対応中
Z-1253-2023 PCHクラス II2023.02.01BioFire Diagnostics, LLC

Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.

対応中
Z-1335-2022 PCHクラス II2022.06.17Luminex Corporation

There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

対応中
Z-2451-2021 PCHクラス II2021.08.03Luminex Corporation

There is potential for false positive results.

対応中
Z-1488-2021 PCHクラス II2021.03.18Luminex Corporation

Potential for a false-negative result for Vibrio parahaemolyticus, Salmonella, and/or Shigella spp when used with the VERIGENE Enteric Pathogens Nucleic Acid Test Kit.

対応中
Z-1451-2020 PCHクラス II2019.08.08BioFire Diagnostics, LLC

Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could potentially lead to more serious health events.

終了
Z-2339-2019 PCHクラス II2019.07.17Becton Dickinson & Co.

The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

終了
Z-2338-2019 PCHクラス II2019.07.17Becton Dickinson & Co.

The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

終了
Z-2330-2019 PCHクラス II2019.07.17Becton Dickinson & Co.

The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

終了
Z-2329-2019 PCHクラス II2019.07.17Becton Dickinson & Co.

The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

終了

510(k)(市販前届出)

全 38 件
No.番号機器名申請者決定日種別概要
1K253722 LIAISON PLEX Gastrointestinal Flex AssayLuminex Corporation2026.05.19通常(Traditional) PDF
2K254032 QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BQIAGEN GmbH2026.03.09通常(Traditional) PDF
3K251721 Xpert GI PanelCepheid2026.01.16通常(Traditional) PDF
4K252329 QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BQIAGEN GmbH2025.10.22通常(Traditional) PDF
5K251993 Panther Fusion GI Expanded Bacterial AssayHologic, Inc.2025.09.25通常(Traditional) PDF
6K251868 Panther Fusion GI Bacterial AssayHologic2025.09.25通常(Traditional) PDF
7K250324 QIAstat-Dx GI Panel 2 Mini BQIAGEN GmbH2025.02.28特別(Special) PDF
8K243885 BIOFIRE FILMARRAY Gastrointestinal (GI) Panel MidBiofire Diagnostics, LLC2025.01.16特別(Special) PDF
9K242877 BioCode Gastrointestinal Pathogen Panel (GPP)Applied BioCode, Inc.2025.01.10通常(Traditional) PDF
10K243813 QIAstat-Dx GI Panel 2 Mini B&VQIAGEN GmbH2025.01.08特別(Special) PDF
11K242367 BIOFIRE FILMARRAY Gastrointestinal (GI) PanelBiofire Diagnostics, LLC2024.11.07通常(Traditional) PDF
12K220062 QIAstat-Dx Gastrointestinal Panel 2QIAGEN GmbH2024.05.31通常(Traditional) PDF
13K232672 EasyScreen Gastrointestinal Parasite Detection KitGenetic Signatures Limited2024.05.29通常(Traditional) PDF
14K230404 BIOFIRE FILMARRAY Gastrointestinal (GI) PanelBiofire Diagnostics, LLC2023.03.16特別(Special) PDF
15K220607 BD MAX Enteric Viral PanelBecton, Dickinson and Company2022.09.21通常(Traditional) PDF
16K220193 BD MAX Enteric Parasite PanelBecton, Dickinson and Company2022.08.19通常(Traditional) PDF
17K191161 xTAG Gastrointestinal Pathogen Panel (GPP)Luminex Molecular Diagnostics, Inc.2019.11.12特別(Special)-
18K191160 xTAG Gastrointestinal Pathogen Panel (GPP), xTAG Data Analysis Software (TDAS GPP)Luminex Molecular Diagnostics, Inc.2019.11.04特別(Special)-
19K182703 EntericBio Dx AssaySerosep , Ltd.2019.06.19通常(Traditional) PDF
20K190585 Biocode Gastrointestinal Pathogen Panel (GPP)Applied BioCode, Inc.2019.06.05通常(Traditional) PDF

出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認

PMA(市販前承認)

この product code の PMA 情報はありません。

不具合報告(MAUDE)トレンド

米国不具合報告検索で詳しく見る

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。

不具合報告データを読み込み中...

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

取得日: 2026.09.28

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。