General Purpose Reagent(PPM)
汎用試薬 この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。
30 件
直近 2026.05.06
16 件
0 件
316 件
分類情報
| Product Code | PPM |
|---|---|
| デバイス名 | General Purpose Reagent 汎用試薬 この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。 |
| デバイスクラス | クラス I |
| 21 CFR 規制番号 | 864.4010 |
| 審査区分 | 届出免除(510(k) Exempt) |
| 医療専門分野 | 病理学 |
| 定義 | A general purpose reagent is a chemical reagent that has general laboratory application, that is used to collect, prepare, and examine specimens from the human body for diagnostic purposes, and that is not labeled or otherwise intended for a specific diagnostic application. It may be either an individual substance, or multiple substances reformulated, which, when combined with or used in conjunction with an appropriate analyte specific reagent (ASR) and other general purpose reagents, is part of a diagnostic test procedure or system constituting a finished in vitro diagnostic (IVD) test. 汎用試薬とは、一般的な実験室用途を有する化学試薬であって、診断目的でヒトの身体から得られた検体を採取、調製、及び検査するために用いられ、かつ特定の診断用途を目的として標識又は意図されていないものをいう。これは単一の物質であってもよく、又は複数の物質を再調製したものであってもよく、適切な分析物特異試薬(ASR)及び他の汎用試薬と組み合わせて用いられる場合、完成した体外診断(IVD)検査を構成する診断検査手順又はシステムの一部となるものである。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。 |
| フラグ | 該当なし |
| AI分析 | 本品は米国においてクラスIに分類される医療機器です。クラスIは相対的にリスクが低い機器を対象とする区分であり、本品は市販前届出(510(k))の対象からも除外される届出免除の扱いとなっています。 米国市場での動向 本製品分野では、510(k)届出が約20件・参入企業約10社に上る一方、直近5年間の新規届出はなく、最新の判定も1987年4月とやや古い時期にとどまっています。不具合報告は累計約310件で、大半が不具合(Malfunction)であり、死亡1件・傷害約10件が報告されています。主な問題としては、結果の不正確・不適切な表示(Incorrect, Inadequate or Imprecise Result or Readings)や試薬・機器成分の汚染(Contamination of Device Ingredient or Reagent/Contamination)が挙げられています。回収は累計約30件で、直近5年では約15件、うち最重篤なクラスIが8件含まれており、最新の回収は2026年3月に開始されています。 この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。 |
関連 product code
同じ 21 CFR 864.4010 に分類される product code(51 件)
| No. | product code | 機器名 | クラス | 専門分野 | 510(k) | PMA | 不具合 | 回収 |
|---|---|---|---|---|---|---|---|---|
| 1 | IFP | Formalin, Neutral Buffered 中性緩衝ホルマリン液 | I | 病理学 | 15 | — | 39 | 2 件 |
| 2 | KEM | Agent, Clearing 清澄化剤 | I | 病理学 | 15 | — | 2 | — |
| 3 | KEO | Formulations, Paraffin, All 組織包埋用パラフィン | I | 病理学 | 11 | — | 1 | 3 件 |
| 4 | KDX | Solution, Decalcifier, Acid Containing 酸性脱灰液 | I | 病理学 | 8 | — | 1 | — |
| 5 | LDZ | Fixative, Alcohol Containing アルコール含有固定液 | I | 病理学 | 8 | — | — | — |
| 6 | IGG | Formaldehyde (Formalin, Formol) ホルムアルデヒド(ホルマリン、フォルモール) | I | 病理学 | 7 | — | 1 | 1 件 |
| 7 | JCC | Ph Buffer pH緩衝液 | I | 病理学 | 7 | — | 164 | 2 件 |
| 8 | JCE | Solution, Isotonic 等張液 | I | 病理学 | 7 | — | 21 | — |
| 9 | KEQ | Media, Mounting, Water Soluble 水溶性標本包埋剤 | I | 病理学 | 7 | — | — | — |
| 10 | LEB | Mounting Media 封入剤 | I | 病理学 | 5 | — | — | — |
| 11 | LDW | Fixative, Acid Containing 酸含有固定液 | I | 病理学 | 4 | — | — | — |
| 12 | LEA | Preservative, Cytological 細胞保存剤 | I | 病理学 | 4 | — | 23 | 4 件 |
機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。
類似の国内医療機器(自動推定)
FDA デバイス分類と JMDN 英語名の名称・定義文の類似度から機械的に推定した候補です。日米の当局が定めた公的な対応関係ではなく、参考情報としてご覧ください。
| No. | 類別 | 一般的名称 | 品目数 | 企業数 | 年間総出荷額 | 成長率 | 平均価格 | 国産比率 |
|---|---|---|---|---|---|---|---|---|
| 1 | 手術台及び治療台 | 汎用診断・処置用テーブルクラスⅠ | 116 | 20 | 10.72 億円/年 | -4.8% | 26.51 万円/個 | 96.5% |
| 2 | 医療用遠心ちんでん器 | 汎用検査室用遠心機クラスⅠ | 120 | 23 | 7.62 億円/年 | +9.5% | 19.67 万円/個 | 89.8% |
| 3 | 手術台及び治療台 | 汎用診療・処置台クラスⅠ | 65 | 30 | 7.89 億円/年 | +7.3% | 105.17 万円/個 | 91.9% |
| 4 | 医療用吸引器 | 再使用可能な汎用吸引チップクラスⅠ | 196 | 92 | 5.22 億円/年 | -23.3% | 79 円/個 | 99.9% |
| 5 | 内臓機能検査用器具 | 汎用血液ガス分析装置クラスⅠ | 43 | 8 | 60.88 億円/年 | -0.8% | 2.75 万円/個 | 3.5% |
FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。
米国での回収・承認・不具合報告
米国回収
全 30 件| 回収番号 | クラス | 回収開始日 | 企業 | 理由 | ステータス |
|---|---|---|---|---|---|
| Z-1984-2026 PPM | クラス II | 2026.03.26 | Immuno-Mycologics, Inc | Neutralization Buffer may contain contaminants | 対応中 |
| Z-0773-2026 PPM | クラス II | 2025.11.03 | GET TESTED INTERNATIONAL AB | Distribution without premarket approval/clearance. | 対応中 |
| Z-0943-2026 PPM | クラス II | 2025.10.29 | Copan Italia | Treatment reagent functionality in respiratory specimens fluidification cannot be guaranteed. | 対応中 |
| Z-0942-2026 PPM | クラス II | 2025.10.29 | Copan Italia | Treatment reagent functionality in respiratory specimens liquification cannot be guaranteed. | 対応中 |
| Z-1993-2025 PPM | クラス II | 2025.05.21 | Diagnostica Stago, Inc. | Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1% | 対応中 |
| Z-1834-2025 PPM | クラス II | 2025.04.24 | Mckesson Medical-Surgical Inc. Corporate Office | transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result. | 対応中 |
| Z-0315-2025 PPM | クラス II | 2024.09.25 | Biocare Medical, LLC | Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated. | 対応中 |
| Z-2266-2024 PPM | クラス II | 2024.05.21 | Ortho-Clinical Diagnostics, Inc. | VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides. | 対応中 |
| Z-1376-2024 PPM | クラス II | 2024.02.07 | Beckman Coulter, Inc. | On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman Coulter has identified that Access Substrate bottles in the identified lots were released from manufacturing without labels. Missing label may cause some patient results reporting delay but will not cause any errorreous results. However, because labs are required to have redundancy such that delay would not significantly impact clinical management and because the defect is easily recognized by the user, the probability of harm is remote. | 対応中 |
| Z-2255-2023 PPM | クラス II | 2023.05.24 | NeuMoDx Molecular Inc | Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results | 対応中 |
| Z-1564-2022 PPM | クラス II | 2022.07.20 | bioMerieux, Inc. | There is a potential of no result leading to possible delayed result due to contamination of extraction reagents. | 対応中 |
| Z-1384-2022 PPM | クラス III(軽微) | 2022.05.20 | Agilent Technologies, Inc. | Label provided for substrate vail contained incorrect expiration date. | 対応中 |
| Z-0824-2022 PPM | クラス III(軽微) | 2022.02.14 | Cytocell Ltd. | Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label. | 終了 |
| Z-0503-2022 PPM | クラス II | 2021.11.22 | Ortho-Clinical Diagnostics, Inc. | Lower than expected results for VITROS¿ Immunodiagnostic Products TSH Reagent Pack when diluted with VITROS¿ Immunodiagnostic Products High Sample Diluent A Reagent Pack Lot 2190. | 対応中 |
| Z-3004-2020 PPM | クラス III(軽微) | 2020.06.24 | DNA Genotek Inc. | Reagents were shipped to customers after the Use by data indicated on its labeling. | 終了 |
| Z-1365-2020 PPM | クラス II | 2020.01.10 | BIOMERIEUX - Grenoble | Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction. | 終了 |
| Z-0443-2019 PPM | クラス II | 2018.10.08 | bioMerieux, Inc. | Some eluates become colored because of remaining hemoglobin in the eluate. | 終了 |
| Z-0218-2019 PPM | クラス II | 2018.09.17 | BioMerieux SA | Data loggers showed that one shipment to the United States experienced low temperatures that may impact performance of this product. | 終了 |
| Z-3016-2018 PPM | クラス I(最重篤) | 2018.08.02 | Ventana Medical Systems Inc | Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur. | 終了 |
| Z-3008-2018 PPM | クラス I(最重篤) | 2018.08.02 | Ventana Medical Systems Inc | Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers | 終了 |
510(k)(市販前届出)
全 16 件| No. | 番号 | 機器名 | 申請者 | 決定日 | 種別 | 概要 |
|---|---|---|---|---|---|---|
| 1 | K871533 | TOMTEC GRAM STAINER DECOLORIZER | Remel Co. | 1987.04.29 | 通常(Traditional) | - |
| 2 | K863550 | POTASSIUM HYDROXIDE 10% W/V | Reagent Laboratory, Inc. | 1986.10.02 | 通常(Traditional) | - |
| 3 | K861890 | 100%, 95%, 80% & 70% DEHYDRANT CUSTOM BLEND NO. 2 | Richard-Allan Medical Ind., Inc. | 1986.06.03 | 通常(Traditional) | - |
| 4 | K850916 | SPRAYFIX | Surgipath Medical Industries, Inc. | 1985.03.25 | 通常(Traditional) | - |
| 5 | K850845 | 0.85% SALINE | Micro Media Laboratories | 1985.03.20 | 通常(Traditional) | - |
| 6 | K850737 | TRANSPORT CULTURE MEDIUM | Roseville Medical Laboratoriesu | 1985.03.20 | 通常(Traditional) | - |
| 7 | K842934 | DECOLORIZING FLUIC, ACID-ALCOHOL 3% | E K Ind., Inc. | 1984.09.12 | 通常(Traditional) | - |
| 8 | K842863 | PRO-SOFT DEHYDRANT | Anatech, Ltd. | 1984.08.21 | 通常(Traditional) | - |
| 9 | K842729 | DECOLORIZING FLUID | E K Ind., Inc. | 1984.08.03 | 通常(Traditional) | - |
| 10 | K842070 | IDD DETERGENT | Interscience Diagnostics, Inc. | 1984.07.02 | 通常(Traditional) | - |
| 11 | K831620 | SODIUM HYPOCHLORITE 3% | Omega Medical Diagnostics | 1983.07.28 | 通常(Traditional) | - |
| 12 | K831619 | SODIUM HYPOCHLORITE 7% | Omega Medical Diagnostics | 1983.07.28 | 通常(Traditional) | - |
| 13 | K822326 | BLENDED DETERGENT-78N 1912-1 | Interscience, Inc. | 1982.10.18 | 通常(Traditional) | - |
| 14 | K822037 | VOLU-SOL TAP WATER SUBSTITUTE | Volu Sol Medical Industries | 1982.08.03 | 通常(Traditional) | - |
| 15 | K811800 | BLUING REAGENT | Richard-Allan Medical Ind., Inc. | 1981.08.12 | 通常(Traditional) | - |
| 16 | K811796 | FORMAL ALCOHOL | Richard-Allan Medical Ind., Inc. | 1981.07.28 | 通常(Traditional) | - |
出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認
PMA(市販前承認)
この product code の PMA 情報はありません。
不具合報告(MAUDE)トレンド
米国不具合報告検索で詳しく見る出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。
報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。
各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。
出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。
取得日: 2026.09.28
各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。