Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems(QLG)
インテグレーテッド持続血糖測定システム(工場出荷時校正済み、自動インスリン投与システム非対応) この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。
10 件
直近 2026.01.21
9 件
0 件
213,817 件
分類情報
| Product Code | QLG |
|---|---|
| デバイス名 | Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems インテグレーテッド持続血糖測定システム(工場出荷時校正済み、自動インスリン投与システム非対応) この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。 |
| デバイスクラス | クラス II |
| 21 CFR 規制番号 | 862.1355 |
| 審査区分 | 510(k)(市販前届出) |
| 医療専門分野 | 臨床化学 |
| 定義 | An integrated continuous glucose monitoring system for replacement of finger stick blood glucose testing for diabetes treatment decisions in persons with diabetes, unless otherwise indicated. The system automatically measures glucose in bodily fluids continuously or frequently for a specified period of time. The system is also designed to reliably and securely transmit glucose measurement data to digitally connected devices and are intended to be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions. The system is not intended to be used with automated insulin dosing (AID) systems. 糖尿病患者において、特に指示がない限り、糖尿病治療の決定を行うための指先穿刺による血糖検査に代わるインテグレーテッド持続血糖測定システムである。本システムは、指定された期間にわたり体液中のグルコースを連続的または頻回に自動測定する。本システムはまた、グルコース測定データをデジタル接続機器に確実かつ安全に送信するよう設計されており、単独で使用される、またはユーザーが治療上の判断のための行動を自ら管理するこれらのデジタル接続機器と併用されることを意図している。本システムは自動インスリン投与(AID)システムと併用することを意図していない。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。 |
| フラグ | 該当なし |
| AI分析 | 本品は米国FDAの規制上クラスIIに分類されており、市販前届出(510(k))の審査区分の対象となっています。510(k)では、既に合法的に市販されている同等の機器(既存機器)との実質的同等性を示すことにより、市販前承認(PMA)よりも簡略化された手続きで承認申請を行うことができます。 米国市場での動向 510(k)としては累計9件が届出されており、申請企業数は2社、直近5年間では5件、最新の判定は2023年3月となっています。不具合報告(MDR)は累計で約161,900件寄せられており、その大半が不具合(Malfunction)に関するもので、死亡は29件、傷害は約41,700件と報告されています。主な問題としては、結果・測定値の不正確・不十分・不明確な表示、測定値取得不能、デバイスの誤表示メッセージが挙げられています。回収は累計10件で、うちクラスI(最重篤)が6件、クラスIIが4件を占め、直近5年間では9件、最新の回収は2025年11月に開始されています。 この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。 |
関連 product code
同じ 21 CFR 862.1355 に分類される product code(6 件)
| No. | product code | 機器名 | クラス | 専門分野 | 510(k) | PMA | 不具合 | 回収 |
|---|---|---|---|---|---|---|---|---|
| 1 | QBJ | Integrated Continuous Glucose Monitoring System, Factory Calibrated 一体型持続グルコースモニタリングシステム(工場校正式) | II | 臨床化学 | 18 | — | 2,268,268 | 23 件 |
| 2 | QDK | Integrated Continuous Glucose Monitoring System For Non-Intensive Diabetes Management 統合型持続血糖測定システム(非強化糖尿病管理用) | II | 臨床化学 | 4 | — | 619 | — |
| 3 | SAF | Integrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-Counter 非強化インスリン療法用一般用一体型持続血糖測定システム | II | 臨床化学 | 4 | — | 2,699 | — |
| 4 | SBH | Integrated Continuous Glucose Monitor For Non-Intensive Glucose Management, Over-The-Counter 非強化インスリン療法用一体型持続グルコースモニタ(一般用) | II | 臨床化学 | 3 | — | 2 | — |
| 5 | QDL | Integrated Continuous Glucose Monitoring System For Professional Retrospective Use 専門家用retrospective使用継続的血糖モニタリングシステム | II | 臨床化学 | 1 | — | — | — |
| 6 | QII | Integrated Continuous Glucose Monitoring System For Professional Directed Retrospective Or Real-Time Use 専門家指導下遡及的又はリアルタイム使用のための統合型持続グルコースモニタリングシステム | II | 臨床化学 | 1 | — | 237 | — |
機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。
類似の国内医療機器(自動推定)
FDA デバイス分類と JMDN 英語名の名称・定義文の類似度から機械的に推定した候補です。日米の当局が定めた公的な対応関係ではなく、参考情報としてご覧ください。
| No. | 類別 | 一般的名称 | 品目数 | 企業数 | 年間総出荷額 | 成長率 | 平均価格 | 国産比率 |
|---|---|---|---|---|---|---|---|---|
| 1 | 体液検査用器具 | グルコースモニタシステムクラスⅢ | 12 | 7 | 436.7 億円/年 | +16.1% | 5,804 円/個 | 0.0% |
| 2 | 尿検査又は糞便検査用器具 | 自己検査用尿糖計クラスⅡ | 1 | 1 | — | — | — | — |
| 3 | 血液検査用器具 | 自己検査用グルコース測定器クラスⅢ | 44 | 13 | 15.69 億円/年 | -11.8% | 2,462 円/個 | 81.1% |
| 4 | 尿検査又は糞便検査用器具 | 電子尿糖計クラスⅡ | — | — | — | — | — | — |
FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。
米国での回収・承認・不具合報告
米国回収
全 10 件| 回収番号 | クラス | 回収開始日 | 企業 | 理由 | ステータス |
|---|---|---|---|---|---|
| Z-1023-2026 QLG | クラス I(最重篤) | 2025.11.24 | Abbott Diabetes Care, Inc. | Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results | 対応中 |
| Z-1022-2026 QLG | クラス I(最重篤) | 2025.11.24 | Abbott Diabetes Care, Inc. | Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results | 対応中 |
| Z-1021-2026 QLG | クラス I(最重篤) | 2025.11.24 | Abbott Diabetes Care, Inc. | Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results | 対応中 |
| Z-1020-2026 QLG | クラス I(最重篤) | 2025.11.24 | Abbott Diabetes Care, Inc. | Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results | 対応中 |
| Z-1997-2025 QLG | クラス II | 2025.05.21 | BIGFOOT BIOMEDICAL | Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision. | 対応中 |
| Z-2692-2024 QLG | クラス I(最重篤) | 2024.07.24 | Abbott Diabetes Care, Inc. | Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death. | 対応中 |
| Z-2414-2023 QLG | クラス II | 2023.07.12 | Abbott Diabetes Care, Inc. | If blood glucose monitoring system users attempt to start a new sensor, when the old sensor wasn't used for the full 14-day wear time, then the reader may display "Incompatible Sensor" message. If users don't have the previous sensor, or FreeStyle Libre 2 app, or the sensor's full wear time hasn't ended, users may be unable to start a new sensor, which could result in no or delayed glucose results | 対応中 |
| Z-1275-2023 QLG | クラス I(最重篤) | 2023.02.13 | Abbott Diabetes Care, Inc. | Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use. | 対応中 |
| Z-0734-2024 QLG | クラス II | 2023.02.09 | Abbott Diabetes Care, Inc. | If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required. | 完了 |
| Z-1931-2021 QLG | クラス II | 2021.03.16 | Cardinal Health Inc. | Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. | 対応中 |
510(k)(市販前届出)
全 9 件| No. | 番号 | 機器名 | 申請者 | 決定日 | 種別 | 概要 |
|---|---|---|---|---|---|---|
| 1 | K222280 | Bigfoot Unity® Diabetes Management System | Bigfoot Biomedical, Inc. | 2023.03.01 | 通常(Traditional) | |
| 2 | K223537 | FreeStyle Libre 2 System, FreeStyle Libre 3 System | Abbott Diabetes Care, Inc. | 2023.02.21 | 通常(Traditional) | |
| 3 | K213996 | FreeStyle Libre 3 Continuous Glucose Monitoring System | Abbott Diabetes Care, Inc. | 2022.05.26 | 通常(Traditional) | |
| 4 | K212132 | FreeStyle Libre 3 Continuous Glucose Monitoring System | Abbott Diabetes Care, Inc. | 2022.05.26 | 通常(Traditional) | |
| 5 | K210943 | FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App) | Abbott Diabetes Care, Inc. | 2021.11.22 | 通常(Traditional) | |
| 6 | K211102 | FreeStyle Libre 2 Flash Glucose Monitoring System | Abbott Diabetes Care, Inc. | 2021.08.11 | 通常(Traditional) | |
| 7 | K201761 | FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App) | Abbott Diabetes Care, Inc. | 2021.07.30 | 通常(Traditional) | |
| 8 | K202145 | Bigfoot Unity Diabetes Management System | Bigfoot Biomedical, Inc. | 2021.05.07 | 通常(Traditional) | |
| 9 | K193371 | FreeStyle Libre 2 Flash Glucose Monitoring System | Abbott Diabetes Care, Inc. | 2020.06.12 | 通常(Traditional) |
出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認
PMA(市販前承認)
この product code の PMA 情報はありません。
不具合報告(MAUDE)トレンド
米国不具合報告検索で詳しく見る出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。
報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。