System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer(NIG)

膵臓癌管理用糖質抗原(CA19-9)検査システム この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

米国回収

19

直近 2024.09.18

510(k)

12

PMA

0

不具合報告

3,190

分類情報

Product Code NIG
デバイス名 System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer

膵臓癌管理用糖質抗原(CA19-9)検査システム この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

デバイスクラス クラス II
21 CFR 規制番号 866.6010
審査区分 510(k)(市販前届出)
医療専門分野 免疫学
定義 An in vitro diagnostic test for the quantitative measurement of the CA 19-9 tumor associated antigen in human serum or plasma. The test is intended as an aid in the management of patients with confirmed pancreatic cancer and serial monitoring of their response to therapy and disease progression. The test should only be used in patients with serum and plasma CA 19-9 values above the cut-off at the time of diagnosis and in conjunction with other clinical methods.

ヒト血清又は血漿中のCA19-9腫瘍関連抗原の定量測定のための体外診断用検査である。本検査は、確定診断された膵臓癌患者の管理、並びに治療反応及び疾患進行の連続的モニタリングの補助として使用することを意図する。本検査は、診断時に血清及び血漿のCA19-9値がカットオフ値を超える患者にのみ使用し、他の臨床的方法と併用すべきである。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。

フラグ
該当なし
AI分析

米国ではクラスIIに分類され、市販前届出(510(k))により審査されます。510(k)は、既に市販されている同等の機器との実質的同等性を示すことにより審査を受ける区分であり、クラスIIIの機器に求められる市販前承認(PMA)に比べて相対的に簡便な手続きとされています。

米国市場での動向

本製品コードでは、510(k)としての累計届出件数は約10件、申請企業数は7社で、直近5年間の届出は1件、最新の判定は2023年8月となっています。不具合報告は累計約3,200件で、その大半がMalfunction(不具合)であり、死亡報告は0件、傷害報告は約50件です。主な問題としてはHigh Test Results、Low Test Results、Non Reproducible Resultsが挙げられています。回収は累計約20件で、クラスIは確認されておらず、クラスIIが約10件、直近5年間の回収は約5件、最新の回収開始は2024年7月です。

この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。

関連 product code

同じ 21 CFR 866.6010 に分類される product code(17 件)

No.product code機器名クラス専門分野510(k)PMA不具合回収
1LTJ

Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers

前立腺癌管理用前立腺特異抗原(PSA)測定キット

II免疫学349065212 件
2DHX

System, Test, Carcinoembryonic Antigen

がん胎児性抗原(CEA)測定用システム

II免疫学28423089 件
3MOI

System, Test, Immunological, Antigen, Tumor

腫瘍抗原免疫学的検査システム

II免疫学2820818 件
4LTK

Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125)

上皮性卵巣腫瘍関連抗原(CA125)検査

II免疫学2662278 件
5LOJ

Kit, Test,Alpha-Fetoprotein For Testicular Cancer

精巣腫瘍用アルファフェトプロテイン測定キット

II免疫学17271107 件
6MSW

System, Test, Thyroglobulin

サイログロブリン測定用試薬

II免疫学138610 件
7MMW

System, Test, Tumor Marker, Monitoring, Bladder

膀胱腫瘍マーカーモニタリング用検査システム

II免疫学961 件
8OIU

Test, Epithelial Ovarian Tumor Associated Antigen (He4)

上皮性卵巣腫瘍関連抗原(HE4)測定キット

II免疫学511
9NCW

System, Test, Her-2/Neu, Monitoring

Her-2/Neuモニタリング検査システム

II免疫学31 件
10OVK

Cytokeratin Fragments 21-1 Eia Kit

サイトケラチンフラグメント21-1測定キット

II免疫学38
11LOQ

Device, General Purpose, Hematology

汎用血液学検査機器

II免疫学1881 件
12NTY

System, Test, Fibrin/Fibrinogen Degradation Products For Monitoring Of Colorectal Cancer

大腸癌モニタリング用フィブリン/フィブリノゲン分解産物測定システム

II免疫学11

機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。

FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。

米国での回収・承認・不具合報告

無料会員 1,000名 突破

回収・承認・不具合報告の明細の閲覧は会員限定です

無料で会員登録する
回収番号クラス回収開始日企業理由ステータス
Z-3104-2024 NIGクラス II2024.07.26Siemens Healthcare Diagnostics, Inc.

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

対応中
Z-3103-2024 NIGクラス II2024.07.26Siemens Healthcare Diagnostics, Inc.

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

対応中
Z-3102-2024 NIGクラス II2024.07.26Siemens Healthcare Diagnostics, Inc.

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

対応中
Z-3101-2024 NIGクラス II2024.07.26Siemens Healthcare Diagnostics, Inc.

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

対応中
Z-3100-2024 NIGクラス II2024.07.26Siemens Healthcare Diagnostics, Inc.

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

対応中
Z-1444-2020 NIGクラス II2020.02.07Roche Diagnostics Operations, Inc.

The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.

終了
Z-2241-2020 NIGクラス II2019.04.30Beckman Coulter Inc.

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

終了
Z-2069-2018 NIGクラス II2018.03.05Tosoh Bioscience Inc

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

終了
Z-2638-2017 NIGクラス II2017.05.09Beckman Coulter Inc.

The Access immunoassays are susceptible to biotin interference.

終了
Z-0613-2017 NIGクラス II2016.10.19Siemens Healthcare Diagnostics, Inc

CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.

終了
Z-0612-2017 NIGクラス II2016.10.19Siemens Healthcare Diagnostics, Inc

CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.

終了
Z-2183-2014 NIGクラス II2014.07.16ORTHO-CLINICAL DIAGNOSTICS

Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

終了
Z-0508-2013 NIGクラス II2012.11.14Siemens Healthcare Diagnostics, Inc

Operating range of the ADVIA Centaur systems is 18C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

終了
Z-1904-2012 NIGクラス II2012.05.30Abbott Laboratories

The six ARCHITECT CA 19-9XR reagent lots are demonstrating a shift up in patient results. The Abbott Controls however do not detect the upward shift and will test within range.

終了
Z-1630-2011 NIG
Z-0781-04 NIG
Z-0489-2012 NIG
Z-0218-05 NIG
Z-0197-05 NIG

510(k)(市販前届出)

全 12 件
No.番号機器名申請者決定日種別概要
1K253528 Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc.2026.08.10通常(Traditional) PDF
2K231525 VITROS Immunodiagnostic Products CA 19-9TM Reagent PackOrtho Clinical Diagnostics2023.08.09通常(Traditional) PDF
3K200997 Lumipulse G CA19-9-NFujirebio Diagnostics,Inc.2020.05.14特別(Special) PDF
4K191973 Lumpipulse G CA19-9-NFujirebio Dianostics, Inc.2019.10.22通常(Traditional) PDF
5K100375 DIMENSION VISTA LOCI CA 19-9 FLEX REAGENT CARTRIDGE, AND DIMENSION LOCI7 CALIBRATOR WITH MODEL(S) K6457, AND KC604Siemens Healthcare Diagnostics, Inc.2011.04.06通常(Traditional) PDF
6K052889 VITROS CA 19-9Fujirebio Diagnostics,Inc.2005.12.20通常(Traditional) PDF
7K052000 ARCHITECT CA 19-9XR ASSAYFujirebio Diagnostics,Inc.2005.10.25通常(Traditional) PDF
8K033038 GI MONITOR AND GI MONITOR CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 387687, 387688Beckman Coulter, Inc.2003.12.29通常(Traditional) PDF
9K031393 BAYER ADVIA CENTAUR CA 19-9 ASSAYBayer Healthcare, LLC2003.06.24通常(Traditional) PDF
10K023240 ST AIA-PACK CA 19-9Tosoh Corp.2002.12.23通常(Traditional)
11K023239 AIA-PACK CA 19-9Tosoh Corp.2002.12.23通常(Traditional)
12K020566 FUJIREBIO DIAGNOSTICS CA 19-9 RIAFujirebio Diagnostics,Inc.2002.05.09通常(Traditional) PDF

出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認

PMA(市販前承認)

この product code の PMA 情報はありません。

不具合報告(MAUDE)トレンド

米国不具合報告検索で詳しく見る

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。

不具合報告データを読み込み中...

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

取得日: 2026.09.07

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。