| Product Code | GEX |
|---|---|
| デバイス名 | Powered Laser Surgical Instrument パワードレーザー手術器械 この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。 |
| デバイスクラス | クラス II |
| 21 CFR 規制番号 | 878.4810 |
| 審査区分 | 510(k)(市販前届出) |
| 医療専門分野 | 一般・形成外科 |
| 定義 | A laser (light amplification by stimulated emission of radiation) based device having coherence, collimated and typically monochromatic radiation. Typically indicated to to cut, destroy, remove or coagulate tissue, generally soft tissue, for general surgical purpose in medical specialties of general and plastic surgery, dermatology/aesthetic, podiatry, otolaryngology (ent), gynecology, neurosurgery, orthopedics (soft tissue), dental and oral surgery, and dentistry. The classification regulation 21 cfr 878.4810 describes a device that is carbon dioxide or argon laser intended to cut, destroy, remove or coagulate tissue by the light. レーザー(放射の誘導放出による光増幅)に基づく機器であり、可干渉性、平行光及び通常は単色性を有する放射を有する。一般外科及び形成外科、皮膚科/美容、足病科、耳鼻咽喉科、婦人科、脳神経外科、整形外科(軟部組織)、歯科及び口腔外科、並びに歯科の各医療専門分野において、一般的な外科的目的のために、組織(一般に軟部組織)を切開、破壊、除去又は凝固させる目的で通常用いられる。分類規則21 CFR 878.4810は、光により組織を切開、破壊、除去又は凝固させることを目的とした炭酸ガスレーザー又はアルゴンレーザーである機器を記述している。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。 |
| フラグ | 該当なし |
| AI分析 | 本機器は米国FDAの規制上クラスIIに分類され、市販前届出(510(k))の対象となっています。クラスIIは中程度のリスクを有する医療機器区分であり、既存の同等な機器との実質的同等性を示すことによって市販が認められる仕組みとなっています。分類規則21 CFR 878.4810に基づき、本製品コードは光を用いて組織を切開、破壊、除去又は凝固させることを目的とした炭酸ガスレーザー又はアルゴンレーザー機器を対象としています。 米国市場での動向 本製品コードは510(k)による届出が累計約2,900件、参入企業数は約790社に上り、直近5年間でも約340件の届出があるなど活発な市場となっています。不具合報告は累計約44,000件で、その大半が不具合(Malfunction、約34,300件)であり、傷害報告は約8,800件、死亡報告は105件となっており、主な問題として破損(Break)、吸引力低下(Decrease in Suction)、機器不良(Defective Device)が挙げられています。回収は累計約200件で、クラスIが2件、クラスIIが約150件、クラスIIIが約2件確認されており、直近5年間では約40件の回収があり、最新の回収は2026年5月に開始されています。 この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。 |
関連 product code
同じ 21 CFR 878.4810 に分類される product code(18 件)
| No. | product code | 機器名 | クラス | 専門分野 | 510(k) | PMA | 不具合 | 回収 |
|---|---|---|---|---|---|---|---|---|
| 1 | OHS | Light Based Over The Counter Wrinkle Reduction 光線式一般用しわ改善機器 | II | 一般・形成外科 | 163 | — | 107 | 4 件 |
| 2 | OHT | Light Based Over-The-Counter Hair Removal 一般用光線脱毛器 | II | 一般・形成外科 | 163 | — | 34 | — |
| 3 | ONF | Powered Light Based Non-Laser Surgical Instrument With Thermal Effect 熱作用式非レーザー光線治療器(電動式) | II | 一般・形成外科 | 74 | — | 33 | — |
| 4 | LNK | Laser For Gastro-Urology Use 消化器・泌尿器用レーザ | II | 一般・形成外科 | 60 | — | 186 | — |
| 5 | OLP | Over-The-Counter Powered Light Based Laser For Acne 家庭用ざ瘡治療光線(レーザー)照射器 | II | 一般・形成外科 | 51 | — | 51 | 4 件 |
| 6 | ONG | Powered Laser Surgical Instrument With MicrobeamFractional Output マイクロビームフラクショナル出力式電動レーザー手術器 | II | 一般・形成外科 | 35 | — | 153 | 4 件 |
| 7 | NUV | Massager, Vacuum, Light Induced Heating 光誘導加熱吸引式マッサージ器 | II | 一般・形成外科 | 20 | — | 10 | — |
| 8 | PDZ | Lasers For Temporary Increase Of Clear Nail In Patients With Onychomycosis 爪白癬用clear nail一時的増加レーザ | II | 一般・形成外科 | 19 | — | 3 | 1 件 |
| 9 | NVK | Laser, Dental, Soft Tissue 歯科用軟組織レーザー | II | 一般・形成外科 | 17 | — | 5 | 2 件 |
| 10 | ONO | Neurosurgical Laser With Mr Thermography MR体温計測機能付き神経外科用レーザー | II | 一般・形成外科 | 8 | — | 1 | 1 件 |
| 11 | ONE | Powered Light Based Non-Laser Surgical Instrument 電動光線式非レーザー手術器械 | II | 一般・形成外科 | 7 | — | 3 | — |
| 12 | OYW | Laser, Cellulite Appearance セルライト外観改善用レーザ | II | 一般・形成外科 | 2 | — | 8 | — |
機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。
類似の国内医療機器(自動推定)
FDA デバイス分類と JMDN 英語名の名称・定義文の類似度から機械的に推定した候補です。日米の当局が定めた公的な対応関係ではなく、参考情報としてご覧ください。
| No. | 類別 | 一般的名称 | 品目数 | 企業数 | 年間総出荷額 | 成長率 | 平均価格 | 国産比率 |
|---|---|---|---|---|---|---|---|---|
| 1 | 医療用のこぎり | 電動式手術用のこぎりクラスⅡ | — | — | — | — | — | — |
| 2 | 電気手術器 | レーザー処置用能動器具クラスⅢ | 2 | 3 | — | — | — | — |
| 3 | 整形用機械器具 | 電動式骨手術器械クラスⅡ | 121 | 44 | 106.5 億円/年 | +13.5% | 15.81 万円/個 | 85.9% |
| 4 | 整形用機械器具 | 歯科用インプラント手術器具クラスⅠ | 650 | 78 | 18.9 億円/年 | +12.3% | 1.31 万円/個 | 25.5% |
| 5 | 視力補正用レンズ | 眼科手術用レーザーレンズクラスⅡ | 3 | 3 | 0.41 億円/年 | -24.6% | 3.98 万円/個 | 0.0% |
FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。
米国での回収・承認・不具合報告
米国回収
全 200 件| 回収番号 | クラス | 回収開始日 | 企業 | 理由 | ステータス |
|---|---|---|---|---|---|
| Z-2554-2026 GEX | クラス II | 2026.05.05 | IPG Medical Corporation | Software anomaly that causes a false display of error code 5018. | 対応中 |
| Z-1695-2026 GEX | クラス II | 2026.02.27 | Olympus Corporation of the Americas | Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. | 対応中 |
| Z-1694-2026 GEX | クラス II | 2026.02.27 | Olympus Corporation of the Americas | Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. | 対応中 |
| Z-1558-2026 GEX | クラス II | 2026.01.27 | Lutronic Corporation | Reports of devices sparking/popping and potentially burning patients. | 対応中 |
| Z-1130-2026 GEX | クラス II | 2025.12.15 | LEASEIR TECHNOLOGIES SLU | The console label for affected devices is missing the "DANGER" symbol. | 完了 |
| Z-2344-2025 GEX | クラス II | 2025.07.22 | LUMENIS, LTD. | The potential for unsterilized product within finished product labeled as sterile. | 対応中 |
| Z-1923-2025 GEX | クラス II | 2025.05.06 | Olympus Corporation of the Americas | Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents. | 対応中 |
| Z-1922-2025 GEX | クラス II | 2025.05.06 | Olympus Corporation of the Americas | Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents. | 対応中 |
| Z-1921-2025 GEX | クラス II | 2025.05.06 | Olympus Corporation of the Americas | Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents. | 対応中 |
| Z-1920-2025 GEX | クラス II | 2025.05.06 | Olympus Corporation of the Americas | Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents. | 対応中 |
| Z-1715-2025 GEX | クラス II | 2025.04.04 | Olympus Corporation of the Americas | Preset treatment parameters are not consistently being used in accordance with the IFU. | 対応中 |
| Z-1290-2025 GEX | クラス II | 2025.01.23 | LUMENIS, LTD. | The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on. | 対応中 |
| Z-0556-2025 GEX | クラス II | 2024.08.23 | WINGDERM ELECTRO-OPTICS LTD. | All model WLA-01 systems sold to US customers are found to have labeling non-compliances | 対応中 |
| Z-0135-2025 GEX | クラス II | 2024.08.13 | Olympus Corporation of the Americas | The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy. | 対応中 |
| Z-0134-2025 GEX | クラス II | 2024.08.13 | Olympus Corporation of the Americas | The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy. | 対応中 |
| Z-2506-2024 GEX | クラス II | 2024.05.31 | Olympus Corporation of the Americas | Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure. | 対応中 |
| Z-2505-2024 GEX | クラス II | 2024.05.31 | Olympus Corporation of the Americas | Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure. | 対応中 |
| Z-2504-2024 GEX | クラス II | 2024.05.31 | Olympus Corporation of the Americas | Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure. | 対応中 |
| Z-1809-2024 GEX | クラス II | 2024.04.01 | Karl Storz Endoscopy | Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection. | 対応中 |
| Z-1385-2024 GEX | クラス II | 2024.01.29 | Olympus Corporation of the Americas | A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations. | 対応中 |
510(k)(市販前届出)
全 2,880 件| No. | 番号 | 機器名 | 申請者 | 決定日 | 種別 | 概要 |
|---|---|---|---|---|---|---|
| 1 | K262426 | Medical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75) | Wuhan Pioon Technology Co., Ltd. | 2026.09.10 | 通常(Traditional) | |
| 2 | K261562 | Diode Laser Therapy System | Sanhe LEFIS Electronics Co., Ltd. | 2026.09.10 | 通常(Traditional) | |
| 3 | K261589 | 1927nm Thulium Laser System (PZ-DJG85-01, PZ-DJG85-02) | Zhengzhou PZ Laser Slim Technology Co., Ltd. | 2026.09.02 | 通常(Traditional) | |
| 4 | K262269 | CO2 Laser System (MBT- CO2 Laser) | Beijing Mega Beauty Technology Co., Ltd. | 2026.09.01 | 通常(Traditional) | |
| 5 | K261821 | 1565nm Laser Therapy Device (PZ-JG86-01,PZ-JG86-02) | Zhengzhou PZ Laser Slim Technology Co., Ltd. | 2026.08.27 | 通常(Traditional) | |
| 6 | K262070 | Diode Laser Therapy Systems (DL612A) | Hebei Zollvan Medical Technology Co., Ltd. | 2026.08.19 | 通常(Traditional) | |
| 7 | K261964 | ECHOLASER X4 | Elesta S.P.A | 2026.08.19 | 通常(Traditional) | |
| 8 | K261841 | Q Switched Nd:YAG Laser machine (QLQW-01) | Beijing Nubway S&T Co., Ltd. | 2026.08.11 | 通常(Traditional) | |
| 9 | K261221 | SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | Olympus Surgical Technologies America | 2026.07.24 | 通常(Traditional) | |
| 10 | K261738 | Medical Thulium Fiber Laser Systems (UroFiber 150Q) | Rhein Laser Technologies Co., Ltd. | 2026.07.21 | 通常(Traditional) | |
| 11 | K261693 | Diode laser hair removal machine (QDTM-04) | Beijing Nubway S&T Co., Ltd. | 2026.07.17 | 通常(Traditional) | |
| 12 | K261674 | Skin Air Cooling System (QALF-01) | Beijing Nubway S&T Co., Ltd. | 2026.07.16 | 通常(Traditional) | |
| 13 | K261701 | Hemolase Fiber (Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8) | Micro-Energy Medical Technology Co., Ltd. | 2026.06.18 | 特別(Special) | |
| 14 | K261570 | PICOSURE WORKSTATION | Cynosure, LLC | 2026.06.17 | 特別(Special) | |
| 15 | K260690 | LumiGlam Laser System (SHE-LSP601-3) | Beijing Sano Laser S&T Development Co.,Ltd | 2026.06.04 | 通常(Traditional) | - |
| 16 | K260489 | LASEmaR 1500 | Eufoton S.R.L. | 2026.05.29 | 通常(Traditional) | |
| 17 | K260138 | LUNE PureHygiene | Enamel Pure | 2026.05.29 | 通常(Traditional) | |
| 18 | K253245 | Boston 2910 (Boston 2910) | Boston Aesthetics, Inc. | 2026.05.21 | 通常(Traditional) | |
| 19 | K260100 | MOSES Raydar | Boston Scientific Corporation | 2026.05.18 | 通常(Traditional) | |
| 20 | K261214 | PICO SHINING (PICO-K; PICOFY) | Speclipse, Inc. | 2026.05.11 | 特別(Special) |
出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認
PMA(市販前承認)
この product code の PMA 情報はありません。
不具合報告(MAUDE)トレンド
米国不具合報告検索で詳しく見る出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。
報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。
各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。
出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。
取得日: 2026.09.28
各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。