Powered Laser Surgical Instrument(GEX)

パワードレーザー手術器械 この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

米国回収

200 件

直近 2026.07.01

510(k)

2,880 件

PMA

0 件

不具合報告

44,282 件

分類情報

Product Code GEX
デバイス名 Powered Laser Surgical Instrument

パワードレーザー手術器械 この日本語名は AI が生成した参考訳です。正式名称は英語の原名です。

デバイスクラス クラス II
21 CFR 規制番号 878.4810
審査区分 510(k)(市販前届出)
医療専門分野 一般・形成外科
定義 A laser (light amplification by stimulated emission of radiation) based device having coherence, collimated and typically monochromatic radiation. Typically indicated to to cut, destroy, remove or coagulate tissue, generally soft tissue, for general surgical purpose in medical specialties of general and plastic surgery, dermatology/aesthetic, podiatry, otolaryngology (ent), gynecology, neurosurgery, orthopedics (soft tissue), dental and oral surgery, and dentistry. The classification regulation 21 cfr 878.4810 describes a device that is carbon dioxide or argon laser intended to cut, destroy, remove or coagulate tissue by the light.

レーザー(放射の誘導放出による光増幅)に基づく機器であり、可干渉性、平行光及び通常は単色性を有する放射を有する。一般外科及び形成外科、皮膚科/美容、足病科、耳鼻咽喉科、婦人科、脳神経外科、整形外科(軟部組織)、歯科及び口腔外科、並びに歯科の各医療専門分野において、一般的な外科的目的のために、組織(一般に軟部組織)を切開、破壊、除去又は凝固させる目的で通常用いられる。分類規則21 CFR 878.4810は、光により組織を切開、破壊、除去又は凝固させることを目的とした炭酸ガスレーザー又はアルゴンレーザーである機器を記述している。 この日本語訳は AI が生成した参考訳です。正確な内容は英語の原文をご確認ください。

フラグ
該当なし
AI分析

本機器は米国FDAの規制上クラスIIに分類され、市販前届出(510(k))の対象となっています。クラスIIは中程度のリスクを有する医療機器区分であり、既存の同等な機器との実質的同等性を示すことによって市販が認められる仕組みとなっています。分類規則21 CFR 878.4810に基づき、本製品コードは光を用いて組織を切開、破壊、除去又は凝固させることを目的とした炭酸ガスレーザー又はアルゴンレーザー機器を対象としています。

米国市場での動向

本製品コードは510(k)による届出が累計約2,900件、参入企業数は約790社に上り、直近5年間でも約340件の届出があるなど活発な市場となっています。不具合報告は累計約44,000件で、その大半が不具合(Malfunction、約34,300件)であり、傷害報告は約8,800件、死亡報告は105件となっており、主な問題として破損(Break)、吸引力低下(Decrease in Suction)、機器不良(Defective Device)が挙げられています。回収は累計約200件で、クラスIが2件、クラスIIが約150件、クラスIIIが約2件確認されており、直近5年間では約40件の回収があり、最新の回収は2026年5月に開始されています。

この分析は公開データ(openFDA・JMDN 対応情報)をもとに AI が生成した参考情報です。市場データは2026年8月時点の集計に基づきます。

関連 product code

同じ 21 CFR 878.4810 に分類される product code(18 件)

No.product code機器名クラス専門分野510(k)PMA不具合回収
1OHS

Light Based Over The Counter Wrinkle Reduction

光線式一般用しわ改善機器

II一般・形成外科163—1074 件
2OHT

Light Based Over-The-Counter Hair Removal

一般用光線脱毛器

II一般・形成外科163—34—
3ONF

Powered Light Based Non-Laser Surgical Instrument With Thermal Effect

熱作用式非レーザー光線治療器(電動式)

II一般・形成外科74—33—
4LNK

Laser For Gastro-Urology Use

消化器・泌尿器用レーザ

II一般・形成外科60—186—
5OLP

Over-The-Counter Powered Light Based Laser For Acne

家庭用ざ瘡治療光線(レーザー)照射器

II一般・形成外科51—514 件
6ONG

Powered Laser Surgical Instrument With MicrobeamFractional Output

マイクロビームフラクショナル出力式電動レーザー手術器

II一般・形成外科35—1534 件
7NUV

Massager, Vacuum, Light Induced Heating

光誘導加熱吸引式マッサージ器

II一般・形成外科20—10—
8PDZ

Lasers For Temporary Increase Of Clear Nail In Patients With Onychomycosis

爪白癬用clear nail一時的増加レーザ

II一般・形成外科19—31 件
9NVK

Laser, Dental, Soft Tissue

歯科用軟組織レーザー

II一般・形成外科17—52 件
10ONO

Neurosurgical Laser With Mr Thermography

MR体温計測機能付き神経外科用レーザー

II一般・形成外科8—11 件
11ONE

Powered Light Based Non-Laser Surgical Instrument

電動光線式非レーザー手術器械

II一般・形成外科7—3—
12OYW

Laser, Cellulite Appearance

セルライト外観改善用レーザ

II一般・形成外科2—8—

機器名の日本語訳は AI により自動生成したものです。正確には英語原文をご確認ください。

類似の国内医療機器(自動推定)

FDA デバイス分類と JMDN 英語名の名称・定義文の類似度から機械的に推定した候補です。日米の当局が定めた公的な対応関係ではなく、参考情報としてご覧ください。

No.類別一般的名称品目数企業数年間総出荷額成長率平均価格国産比率
1医療用のこぎり電動式手術用のこぎりクラスⅡ——————
2電気手術器レーザー処置用能動器具クラスⅢ23————
3整形用機械器具電動式骨手術器械クラスⅡ12144106.5 億円/年+13.5% 15.81 万円/個85.9%
4整形用機械器具歯科用インプラント手術器具クラスⅠ6507818.9 億円/年+12.3% 1.31 万円/個25.5%
5視力補正用レンズ眼科手術用レーザーレンズクラスⅡ330.41 億円/年-24.6% 3.98 万円/個0.0%

FDA のデバイスクラスは 3 段階(クラス I〜III)、日本の医療機器は 4 段階(クラス I〜IV)で分類されており、 1 対 1 では対応しません。 リスク区分の考え方が異なるため、参考情報としてご利用ください。

米国での回収・承認・不具合報告

無料会員 1,000名 突破

回収・承認・不具合報告の明細の閲覧は会員限定です

無料で会員登録する
回収番号クラス回収開始日企業理由ステータス
Z-2554-2026 GEXクラス II2026.05.05IPG Medical Corporation

Software anomaly that causes a false display of error code 5018.

対応中
Z-1695-2026 GEXクラス II2026.02.27Olympus Corporation of the Americas

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

対応中
Z-1694-2026 GEXクラス II2026.02.27Olympus Corporation of the Americas

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

対応中
Z-1558-2026 GEXクラス II2026.01.27Lutronic Corporation

Reports of devices sparking/popping and potentially burning patients.

対応中
Z-1130-2026 GEXクラス II2025.12.15LEASEIR TECHNOLOGIES SLU

The console label for affected devices is missing the "DANGER" symbol.

完了
Z-2344-2025 GEXクラス II2025.07.22LUMENIS, LTD.

The potential for unsterilized product within finished product labeled as sterile.

対応中
Z-1923-2025 GEXクラス II2025.05.06Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

対応中
Z-1922-2025 GEXクラス II2025.05.06Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

対応中
Z-1921-2025 GEXクラス II2025.05.06Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

対応中
Z-1920-2025 GEXクラス II2025.05.06Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

対応中
Z-1715-2025 GEXクラス II2025.04.04Olympus Corporation of the Americas

Preset treatment parameters are not consistently being used in accordance with the IFU.

対応中
Z-1290-2025 GEXクラス II2025.01.23LUMENIS, LTD.

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

対応中
Z-0556-2025 GEXクラス II2024.08.23WINGDERM ELECTRO-OPTICS LTD.

All model WLA-01 systems sold to US customers are found to have labeling non-compliances

対応中
Z-0135-2025 GEXクラス II2024.08.13Olympus Corporation of the Americas

The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

対応中
Z-0134-2025 GEXクラス II2024.08.13Olympus Corporation of the Americas

The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

対応中
Z-2506-2024 GEXクラス II2024.05.31Olympus Corporation of the Americas

Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

対応中
Z-2505-2024 GEXクラス II2024.05.31Olympus Corporation of the Americas

Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

対応中
Z-2504-2024 GEXクラス II2024.05.31Olympus Corporation of the Americas

Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

対応中
Z-1809-2024 GEXクラス II2024.04.01Karl Storz Endoscopy

Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.

対応中
Z-1385-2024 GEXクラス II2024.01.29Olympus Corporation of the Americas

A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations.

対応中

510(k)(市販前届出)

全 2,880 件
No.番号機器名申請者決定日種別概要
1K262426 Medical Diode Laser (S1Pro+, S1Pro K15, S1Pro N6, S1Pro N75)Wuhan Pioon Technology Co., Ltd.2026.09.10通常(Traditional) PDF
2K261562 Diode Laser Therapy SystemSanhe LEFIS Electronics Co., Ltd.2026.09.10通常(Traditional) PDF
3K261589 1927nm Thulium Laser System (PZ-DJG85-01, PZ-DJG85-02)Zhengzhou PZ Laser Slim Technology Co., Ltd.2026.09.02通常(Traditional) PDF
4K262269 CO2 Laser System (MBT- CO2 Laser)Beijing Mega Beauty Technology Co., Ltd.2026.09.01通常(Traditional) PDF
5K261821 1565nm Laser Therapy Device (PZ-JG86-01,PZ-JG86-02)Zhengzhou PZ Laser Slim Technology Co., Ltd.2026.08.27通常(Traditional) PDF
6K262070 Diode Laser Therapy Systems (DL612A)Hebei Zollvan Medical Technology Co., Ltd.2026.08.19通常(Traditional) PDF
7K261964 ECHOLASER X4Elesta S.P.A2026.08.19通常(Traditional) PDF
8K261841 Q Switched Nd:YAG Laser machine (QLQW-01)Beijing Nubway S&T Co., Ltd.2026.08.11通常(Traditional) PDF
9K261221 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)Olympus Surgical Technologies America2026.07.24通常(Traditional) PDF
10K261738 Medical Thulium Fiber Laser Systems (UroFiber 150Q)Rhein Laser Technologies Co., Ltd.2026.07.21通常(Traditional) PDF
11K261693 Diode laser hair removal machine (QDTM-04)Beijing Nubway S&T Co., Ltd.2026.07.17通常(Traditional) PDF
12K261674 Skin Air Cooling System (QALF-01)Beijing Nubway S&T Co., Ltd.2026.07.16通常(Traditional) PDF
13K261701 Hemolase Fiber (Hemo-R-0.40-2.5/Hemo-R-0.60-2.5/Hemo-R-0.40-1.8/Hemo-R-0.60-1.8)Micro-Energy Medical Technology Co., Ltd.2026.06.18特別(Special) PDF
14K261570 PICOSURE WORKSTATIONCynosure, LLC2026.06.17特別(Special) PDF
15K260690 LumiGlam Laser System (SHE-LSP601-3)Beijing Sano Laser S&T Development Co.,Ltd2026.06.04通常(Traditional)-
16K260489 LASEmaR 1500Eufoton S.R.L.2026.05.29通常(Traditional) PDF
17K260138 LUNE PureHygieneEnamel Pure2026.05.29通常(Traditional) PDF
18K253245 Boston 2910 (Boston 2910)Boston Aesthetics, Inc.2026.05.21通常(Traditional) PDF
19K260100 MOSES Raydar™Boston Scientific Corporation2026.05.18通常(Traditional) PDF
20K261214 PICO SHINING (PICO-K; PICOFY)Speclipse, Inc.2026.05.11特別(Special) PDF

出典の openFDA は各 510(k) の主要(primary)product code のみを収録しているため、副次コードを含む FDA 公式データベースの検索結果とは件数が異なる場合があります。 FDA 510(k) データベースで確認

PMA(市販前承認)

この product code の PMA 情報はありません。

不具合報告(MAUDE)トレンド

米国不具合報告検索で詳しく見る

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

報告の存在は、機器と事象の因果関係を意味しません。報告件数を機器間の不具合発生率の比較に用いることはできません。

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。

不具合報告データを読み込み中...

出典: openFDA(U.S. FDA) のデータをそのまま掲載しています。内容の正確性・最新性は保証されません。FDA が本サイトの内容を承認・保証するものではありません。

取得日: 2026.09.28

各データセット(510(k)・PMA・MAUDE・回収・Warning Letter)の解説は 米国FDA情報の概要 を参照してください。